Safety requirements for electrical equipment for measurement, control and laboratory use - Part 2-061: Particular requirements for laboratory atomic spectrometers with thermal atomization and ionization (IEC 61010-2-061:2015)

This part of IEC 61010 applies to electrically powered laboratory atomic spectrometers with
thermal atomization.
NOTE 1 Examples include atomic absorption spectrometers, emission flame photometers, atomic fluorescence
spectrophotometers, inductively coupled plasma spectrometers, microwave coupled plasma spectrometers and
mass spectrometers, all with thermal atomization and ionization (including tubing and connectors which are
provided by the manufacturer for connection to external supplies).
NOTE 2 If all or part of the equipment falls within the scope of one or more other part 2 standards of IEC 61010
as well as within the scope of this standard, consideration is to be given to those other part 2 standards.

Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel- und Laborgeräte - Teil 2-061: Besondere Anforderungen an Labor-Atomspektrometer mit thermischer Atomisierung und Ionisation (IEC 61010-2-061:2015)

Règles de sécurité pour appareils électriques de mesurage, de régulation et de laboratoire - Partie 2-061: Prescriptions particulières pour spectromètres de laboratoire avec vaporisation et ionisation thermiques (IEC 61010-2-061:2015)

L'IEC 61010-2-061:2015 s'applique aux spectromètres atomiques de laboratoire équipés d'une alimentation électrique et d'une vaporisation thermique. Elle a le statut d'une publication groupée de sécurité conformément au Guide IEC 104. Cette troisième édition annule et remplace la deuxième édition parue en 2003. Cette édition constitue une révision technique et inclut les modifications suivantes par rapport à la deuxième édition:
- exclusion des appareils dont la taille et le poids rendent peu probables les mouvements involontaires, de l'essai de chute à l'Article 8,
- reformulation des notes selon les Directives ISO/IEC.

Varnostne zahteve za električno opremo za meritve, nadzor in laboratorijsko uporabo - 2-061. del: Posebne zahteve za laboratorijske atomske spektrometre s termično atomizacijo in ionizacijo (IEC 61010-2-061:2015)

Ta del standarda IEC 61010 velja za električno napajane laboratorijske atomske spektrometre s termično atomizacijo.
OPOMBA 1 Primeri vključujejo atomske absorpcijske spektrometre, emisijske plamenske fotometre, atomske fluorescenčne spektrofotometre, spektrometre z induktivno sklopljeno plazmo, spektrometre z mikrovalovno sklopljeno plazmo in masne spektrometre, vse s termalno atomizacijo in ionizacijo (vključno z ocevjem in priključki, ki jih zagotovi proizvajalec za priključitev na zunanja sredstva).
OPOMBA 2 Če vsa oprema ali njen del sodi v okvir enega ali več drugih delov standarda IEC 61010 in v okvir tega standarda, bo treba upoštevati tudi zahteve teh drugih delov standardov.

General Information

Status
Published
Publication Date
28-Jun-2015
Current Stage
6060 - National Implementation/Publication (Adopted Project)
Start Date
14-May-2015
Due Date
19-Jul-2015
Completion Date
29-Jun-2015

Relations

Buy Standard

Standard
EN 61010-2-061:2015
English language
18 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)

SLOVENSKI STANDARD
SIST EN 61010-2-061:2015
01-september-2015
9DUQRVWQH]DKWHYH]DHOHNWULþQRRSUHPR]DPHULWYHQDG]RULQODERUDWRULMVNR
XSRUDERGHO3RVHEQH]DKWHYH]DODERUDWRULMVNHDWRPVNHVSHNWURPHWUHV
WHUPLþQRDWRPL]DFLMRLQLRQL]DFLMR ,(&
Safety requirements for electrical equipment for measurement, control and laboratory
use - Part 2-061: Particular requirements for laboratory atomic spectrometers with
thermal atomization and ionization (IEC 61010-2-061:2015)
Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel- und Laborgeräte - Teil 2
-061: Besondere Anforderungen an Labor-Atomspektrometer mit thermischer
Atomisierung und Ionisation (IEC 61010-2-061:2015)
Règles de sécurité pour appareils électriques de mesurage, de régulation et de
laboratoire - Partie 2-061: Prescriptions particulières pour spectromètres de laboratoire
avec vaporisation et ionisation thermiques (IEC 61010-2-061:2015)
Ta slovenski standard je istoveten z: EN 61010-2-061:2015
ICS:
19.080 (OHNWULþQRLQHOHNWURQVNR Electrical and electronic
SUHVNXãDQMH testing
71.040.10 Kemijski laboratoriji. Chemical laboratories.
Laboratorijska oprema Laboratory equipment
SIST EN 61010-2-061:2015 en
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------

SIST EN 61010-2-061:2015

---------------------- Page: 2 ----------------------

SIST EN 61010-2-061:2015


EUROPEAN STANDARD EN 61010-2-061

NORME EUROPÉENNE

EUROPÄISCHE NORM
April 2015
ICS 19.080; 71.040.20 Supersedes EN 61010-2-061:2003
English Version
Safety requirements for electrical equipment for measurement,
control and laboratory use - Part 2-061: Particular requirements
for laboratory atomic spectrometers with thermal atomization and
ionization
(IEC 61010-2-061:2015)
Règles de sécurité pour appareils électriques de mesurage, Sicherheitsbestimmungen für elektrische Mess-, Steuer-,
de régulation et de laboratoire - Partie 2-061: Exigences Regel- und Laborgeräte - Teil 2-061: Besondere
particulières pour spectromètres atomiques de laboratoire Anforderungen an Labor-Atomspektrometer mit thermischer
avec vaporisation et ionisation thermiques Atomisierung und Ionisation
(IEC 61010-2-061:2015) (IEC 61010-2-061:2015)
This European Standard was approved by CENELEC on 2015-04-14. CENELEC members are bound to comply with the CEN/CENELEC
Internal Regulations which stipulate the conditions for giving this European Standard the status of a national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC
Management Centre or to any CENELEC member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation
under the responsibility of a CENELEC member into its own language and notified to the CEN-CENELEC Management Centre has the
same status as the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,
Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia,
Lithuania, Luxembourg, Malta, the Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Turkey and the United Kingdom.


European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Avenue Marnix 17, B-1000 Brussels
© 2015 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.
 Ref. No. EN 61010-2-061:2015 E

---------------------- Page: 3 ----------------------

SIST EN 61010-2-061:2015
EN 61010-2-061:2015 - 2 -
Foreword
The text of document 66/553/FDIS, future edition 3 of IEC 61010-2-061, prepared by
IEC/TC 66 "Safety of measuring, control and laboratory equipment" was submitted to the
IEC-CENELEC parallel vote and approved by CENELEC as EN 61010-2-061:2015.

The following dates are fixed:
(dop) 2016-01-14
• latest date by which the document has to be
implemented at national level by
publication of an identical national
standard or by endorsement
• latest date by which the national (dow) 2018-04-14
standards conflicting with the
document have to be withdrawn

This document supersedes EN 61010-2-061:2003.

Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CENELEC [and/or CEN] shall not be held responsible for identifying any or all such
patent rights.

This standard covers the Principle Elements of the Safety Objectives for Electrical Equipment
Designed for Use within Certain Voltage Limits (LVD - 2006/95/EC).

Endorsement notice
The text of the International Standard IEC 61010-2-061:2015 was approved by CENELEC as a
European Standard without any modification.

---------------------- Page: 4 ----------------------

SIST EN 61010-2-061:2015
- 3 - EN 61010-2-061:2015

Annex ZA
(normative)
Normative references to international publications
with their corresponding European publications
The following documents, in whole or in part, are normatively referenced in this document and are
indispensable for its application. For dated references, only the edition cited applies. For undated
references, the latest edition of the referenced document (including any amendments) applies.
NOTE 1 When an International Publication has been modified by common modifications, indicated by (mod), the relevant

EN/HD applies.

NOTE 2 Up-to-date information on the latest versions of the European Standards listed in this annex is available here:
www.cenelec.eu.

Annex ZA of Part 1 applies.

---------------------- Page: 5 ----------------------

SIST EN 61010-2-061:2015

---------------------- Page: 6 ----------------------

SIST EN 61010-2-061:2015



IEC 61010-2-061

®


Edition 3.0 2015-03




INTERNATIONAL



STANDARD




NORME



INTERNATIONALE




GROUP SAFETY PUBLICATION

PUBLICATION GROUPÉE DE SÉCURITÉ






Safety requirements for electrical equipment for measurement, control, and

laboratory use –

Part 2-061: Particular requirements for laboratory atomic spectrometers with


thermal atomization and ionization



Règles de sécurité pour appareils électriques de mesurage, de régulation et de


laboratoire –

Partie 2-061: Exigences particulières pour spectromètres atomiques de

laboratoire avec vaporisation et ionisation thermique










INTERNATIONAL

ELECTROTECHNICAL

COMMISSION


COMMISSION

ELECTROTECHNIQUE


INTERNATIONALE




ICS 19.080; 71.040.20 ISBN 978-2-8322-2302-4



Warning! Make sure that you obtained this publication from an authorized distributor.

Attention! Veuillez vous assurer que vous avez obtenu cette publication via un distributeur agréé.

® Registered trademark of the International Electrotechnical Commission
Marque déposée de la Commission Electrotechnique Internationale

---------------------- Page: 7 ----------------------

SIST EN 61010-2-061:2015
– 2 – IEC 61010-2-061:2015 © IEC 2015
CONTENTS
FOREWORD . 3
1 Scope and object . 5
2 Normative references . 5
3 Terms and Definitions . 5
4 Tests . 6
5 Marking and documentation . 6
6 Protection against electric shock . 9
7 Protection against mechanical hazards . 9
8 Resistance to mechanical stresses . 9
9 Protection against the spread of fire . 10
10 Equipment temperature limits and resistance to heat . 10
11 Protection against HAZARDS from fluids . 10
12 Protection against radiation, including laser sources, and against sonic and
ultrasonic pressure . 11
13 Protection against liberated gases and substances, explosion and implosion . 11
14 Components and subassemblies . 13
15 Protection by interlocks . 13
16 HAZARDS resulting from application . 14
17 RISK Assessment . 14
Annexes . 15
Annex F (normative) Routine tests . 15
Annex H (informative) Index of defined terms . 15
Bibliography . 15

Figure 101 – Example of a GAS LOCK . 12

Table 101 – Time-temperature conditions . 9

---------------------- Page: 8 ----------------------

SIST EN 61010-2-061:2015
IEC 61010-2-061:2015 © IEC 2015 – 3 –
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________

SAFETY REQUIREMENTS FOR ELECTRICAL EQUIPMENT
FOR MEASUREMENT, CONTROL, AND LABORATORY USE –

Part 2-061: Particular requirements for laboratory atomic
spectrometers with thermal atomization and ionization

FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International Standard IEC 61010-2-061 has been prepared by IEC technical committee 66:
Safety of measuring, control and laboratory equipment.
It has the status of a group safety publication in accordance with IEC Guide 104.
This third edition cancels and replaces the second edition published in 2003. It constitutes a
technical revision and includes the following change from the second edition:
• exclusion of equipment, whose size and weight make unintentional movement unlikely,
from the drop test in Clause 8.
• notes have been re-phrased according to ISO/IEC Directives.

---------------------- Page: 9 ----------------------

SIST EN 61010-2-061:2015
– 4 – IEC 61010-2-061:2015 © IEC 2015
The text of this standard is based on the following documents:
FDIS Report on voting
66/553/FDIS 66/568/RVD

Full information on the voting for the approval of this standard can be found in the report on
voting indicated in the above table.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
A list of all parts of the IEC 61010 series, under the general title: Safety requirements for
electrical equipment for measurement, control, and laboratory use, may be found on the IEC
website.
This Part 2-061 is intended to be used in conjunction with IEC 61010-1. It was established
on the basis of the third edition (2010).
This Part 2-061 supplements or modifies the corresponding clauses in IEC 61010-1 so as to
convert that publication into the IEC standard: Safety requirements for laboratory atomic
spectrometers with thermal atomization and ionization.
Where a particular subclause of Part 1 is not mentioned in this part 2, that subclause applies
as far as is reasonable. Where this part states “addition”, “modification” or “replacement”, or
“deletion”, the relevant requirement, test specification or note in Part 1 should be adapted
accordingly.
In this standard:
1) the following print types are used:
– requirements: in roman type;
– NOTES: in small roman type;
– conformity and test: in italic type;
– terms used throughout this standard which have been defined in clause 3: SMALL ROMAN
CAPITALS;
2) subclauses, figures, tables and notes which are additional to those in Part 1 are numbered
starting from 101. The additional annexes are lettered starting from AA.
The committee has decided that the contents of this publication will remain unchanged until
the stability date indicated on the IEC web site under "http://webstore.iec.ch" in the data
related to the specific publication. At this date, the publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.

---------------------- Page: 10 ----------------------

SIST EN 61010-2-061:2015
IEC 61010-2-061:2015 © IEC 2015 – 5 –
SAFETY REQUIREMENTS FOR ELECTRICAL EQUIPMENT
FOR MEASUREMENT, CONTROL, AND LABORATORY USE –

Part 2-061: Particular requirements for laboratory atomic
spectrometers with thermal atomization and ionization



1 Scope and object
This clause of Part 1 is applicable except as follows:
1.1 Scope
1.1.1 Equipment included in scope
Replacement:
Replace the text by the following:
This part of IEC 61010 applies to electrically powered laboratory atomic spectrometers with
thermal atomization.
NOTE 1 Examples include atomic absorption spectrometers, emission flame photometers, atomic fluorescence
spectrophotometers, inductively coupled plasma spectrometers, microwave coupled plasma spectrometers and
mass spectrometers, all with thermal atomization and ionization (including tubing and connectors which are
provided by the manufacturer for connection to external supplies).
NOTE 2 If all or part of the equipment falls within the scope of one or more other part 2 standards of IEC 61010
as well as within the scope of this standard, consideration is to be given to those other part 2 standards.
1.1.2 Equipment excluded from scope
Addition:
Add as the first paragraph:
This standard does not apply to thermal atomization detectors (flame ionization detectors)
used in gas chromatography.
2 Normative references
This clause of Part 1 is applicable.
3 Terms and Definitions
This clause of Part 1 is applicable except as follows:
Additions:
Add the following definitions:
3.2.101
SPRAY CHAMBER
chamber in which droplets of sample in aerosol are allowed to separate so that the droplets of
necessary size can be passed onward to the burner, with the remainder draining to waste

---------------------- Page: 11 ----------------------

SIST EN 61010-2-061:2015
– 6 – IEC 61010-2-061:2015 © IEC 2015
3.2.102
GAS LOCK
device to allow drainage of waste sample liquid, and to prevent unintentional escape of gas
from the SPRAY CHAMBER through its drain outlet
Note 1 to entry: See for example Figure 101.
3.5.101
FLASH-BACK
event during which the flame travels back through the burner with the result that the gas in
the mixing chamber is caused to ignite.
4 Tests
This clause of Part 1 is applicable except as follows:
Addition:
Add the following subclauses:
4.4.2.101 Sampling probe tip
Any system designed to withdraw a sampling probe tip after sampling has been completed
shall be overridden, so as to leave the tip in its most exposed position when a sample vessel
is removed.
Exceptions:
The withdrawal system need not be overridden if the sampling probe:
a) cannot cause a HAZARD to the OPERATOR when it is exposed;
b) is designed in such a manner that
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.